- 880nm LED Light Source.
- Uses Nephelometric principle with 90° detection.
- Selectable turbidity units, including the NTU, FNU and EBC.
Benchtop Turbidity Meter, BEP-TB1000/BEP-TB200L
Description
Description
The desktop turbidimeter has a built-in microprocessor, advanced configuration, and powerful functions. It is a very precise turbidity measuring instrument. It is based on the infrared light source that passes through the optical lens and penetrates the sample liquid.
This benchtop turbidity meter can be used to measure the turbidity of raw or purified water on filtration devices in various places, such as drinking water, various production and industrial waters, and wherever qualified water is required.
Features
- 880nm LED Light Source.
- Uses Nephelometric principle with 90° detection.
- Offers consistent light source for higher accuracy.
- Selectable turbidity units, including the NTU, FNU and EBC.
- Settable parameters, including user ID, number of calibration
points, data and time, etc.
BEP-TB1000
- 1-7 points calibration using Formazin standards or zero water.
- Autoranging from 0-1000 NTUs.
- Data capacity of up to 2000 sets (GLP-compliant).
- Data stored can be transferred into the computer by USB or RS-232 communication interface.
BEP-TB200L
- 1-5 points calibration using Formazin standards or zero water.
- Autoranging from 0-200 NTUs.
Specifications
Model | BEP-TB1000 | BEP-TB200L | |
Measuring Unit | Range | (0~20.00) NTU,
(20.0~200.0) NTU (200~1000) NTU |
(0~20.00) NTU,
(20.0~200.0) NTU |
Accuracy | ≤±6% | ≤±8% | |
Repeatability | ≤0.5% | ≤1.0% | |
Fluctuation | ≤±0.5%FS/30min | ≤±1.0% FS/30min | |
General | Electricity | AC Adapter, 200-240 VAC input | AC Adapter, 200-240 VAC input |
Dimension (mm)& Weight (kg) | 330×250×130, 2.5 | 330×250×130, 2.5 | |
Package Dimension (mm)& G.W.(kg) | 390×320×360,3.8 |