Chat with us, powered by LiveChat

Blood Gas Analyzer, BGA-95-4

  • Test items:pH,pCO2, pO2 ,Hct
  • Sample:whole blood, serum, plasma, dialysate, CSF
  • Analysis time:<90s
  • Sample volume: 95μL Whole blood;50μL Capillary blood

Share the product

Категория:

Описание

Description

This new designed blood gas analyzer has a convenient disposable calibrants cartridge containing flush solution, mixed gas and liquid calibrants, waste container. This new cartridge design can avoid operator biological hazards and environment pollution.

Features

  • Intuitive touch screen

High resolution 10.4’’ touch screen

User-friendly operation interface

Parameter area provides analyzer status at a glance

Online operation and self-diagnostic tutorials

  • Infrared human-detector
  • Flexible and simple calibrants cartridge

High and low volume testing cartridge can be chose(300,200,150,100,50).

30 days available on board

Stored at room temperature

All-in-one design without extra consumables

Cartridge expiration alarm and residue information

  • Auto QC cartridge

Optional Auto QC cartridge containing 3 levels QC design

Auto H/M/L QC once per days, customer can define

30 days available on board

Stored at room temperature

All-in-one design without extra consumables

Cartridge expiration alarm and residue information

Auto-stop analysis sample once failing on QC

  • Reliable and convenient measurement unit

Capillary Sample volume only 50μL

Whole blood volume only 95μL

Multi-combination parameters

Self-made long life maintenance free electrodes

Maximum 90s result from aspiration, less than 120s from sample to sample

Long-lasting fluid system

Preheats reagents and sample

  • Easy to use

The sample adaptor is suitable for both syringe and capillary samples.

The sample probe’s wiper provides convenience, sample integrity and user safety

  • Fast data input

Easy sample and patient ID input

Convenient bar-code calibrants and Auto-QC cartridge replace

Easy bar-code consumables identification

  • Rich data management

RS-232 interface

LIS connection under HL7 protocol

Data backup with USB port

  • Compact, flexible

Size(mm) W*H*D: 400×574×344

Easy move with rolling stand

Nickel-metal hydride battery,3.0Ah

  • Test Menu

pH, pCO2 , pO2 , Hct

Measured Parameters

Analyte Measuring range Unit
Hydrogen ion activity(pH) 6.000―9.000 pH scale
Carbon dioxide 0.667―26.67 kPa
partial pressure(pCO2) 5.0―200.0 mmHg
Oxygen partial pressure(pO2) 0―106.7 kPa
0―800.0 mmHg
Hct 12.0―65.0 %
B.P. 500―800 mmHg

 

Derived(Calculated) Parameters:

pH(TC) pH temperature corrected
pCO2 (TC) pCO2 temperature corrected
pO2 (TC) pO2 temperature corrected
TCO2 0-50mmol/L
HCO3 0-50mmol/L
Beb -25.0-25mmol/L
Beecf -25.0-25mmol/L
SBC 0-50mmol/L
SO2% 40-100.0%
AG Anion gap
RI 0.0-70.0
AaDO2 0-700mmHg
O2SAT Oxygen Saturation
tHb(c) Hemoglobin(Calculated)
Sample Whole blood, Serum, Plasma, Dialysate, CSF
Test items pH,pCO2, pO2 ,Hct
Input parameter Patient temperature

Hemoglobin

FiO

B.P.

Interface Serial Line RS-232

TCP/IP interface

USB port

Integrated Bar-code reader

Analysis time <90s
Sample volume 95μL Whole blood

50μL Capillary blood

Calibration Automatic, programmable1-point and 2-point calibration
Data storage >10000
Display  10.4” TFT touch screen
Battery Nickel-metal hydride battery 3.0Ah
Printer  Build-in Thermal printer70mm
Barcode reader  Up to 14 kinds barcode type
Instrument 400(W)*574(H)*344(D)mm3
CAL PAK 258(W)*180(H)*70(D)mm3
AUTOQC 103(W)*180(H)*70(D)mm3
Weight 30Kg
Work environment 15℃~30℃
Storage environment 5℃~40℃
Electricity 100-240V~50/60Hz

    Get Quotes Now!

    We will usually contact you within 24 hours. You could also call us (+86-531-88982330) during working hours ( 8:30 am to 5:45 pm UTC+8 Mon.~Sat. ) or use the website live chat to get prompt reply.

    Contact Info

    Rm. 2014, Bldg. 3, Ligaoguojihuayuan, No. 1222, West Aoti Road, Lixia District, Jinan, Shandong

    Phone: +86-531-88982330

    Fax: +86-531-88983691

    Web: www.infitek.com