- Extra-large high-definition color capacitive touch screen
- Intelligent maintenance tracking and reminders
- High-precision gas path control system
Description
The GC-906 Gas Chromatograph seamlessly integrates cutting-edge technology with a user-centred design philosophy. Equipped with a technical platform and data processing system, it is used in many laboratories and analytical institutions due to its good performance, wide range of applications and intelligent operation mode.
Features
Compatible Products & Accessories
Specification
GC-906 | |||
Host | Pressure/Flow Control | Supports up to 18 channels of electronic pressure/flow control (EPC) | |
Pressure Control Accuracy | 0.001psi | ||
Number of sample ports supported | 3 | ||
Maximum number of detectors supported | 4 | ||
Temperature Control System | 8-channel independent temperature control system | ||
Smart Maintenance Tracking and Reminders | yes | ||
Column oven | Control temperature | Room temperature+2℃ ~ 450℃ | |
Minimum step value | 0.01℃ | ||
Maximum heating rate | ≥250ºC/min | ||
Cooling time | From 450°C to 50°C in <3.2 minutes | ||
Programmed Temperature Ramp | ≥99 steps 100 Platform | ||
Temperature Stability | Ambient temperature ±1°C, column oven temperature variation <0.01°C | ||
Others | 1. Automatic furnace chamber temperature protection function 2. Dual column box independent temperature control |
||
Injection System | Injection type | Compatible with split/splitless capillary injectors, septum purging packed column injectors, cold column head injectors, programmable temperature vaporization injection ports, valve injection systems, headspace injection systems, thermal desorption injection systems, etc. | |
Maximum operating temperature | 450℃ | ||
Programmatic current increase | ≥ 10 steps | ||
Stem pressure | 0~50psi、0-200psi | ||
Pressure Control Accuracy | 0.001Psi | ||
Maximum bypass ratio | 1:10000 | ||
Flow Setting Range | 0~500ml/min 0~1250ml/min |
||
Flow Setting Accuracy | 0.01ml/min | ||
Carrier Gas Control Mode | Constant pressure, constant flow rate, constant linear velocity, programmed pressure, programmed flow rate, programmed linear velocity, pulse injection | ||
Others | Features an independent split cold trap and a septum purge trap | ||
Detector | Hydrogen Flame Ionization Detector (FID) | Maximum operating temperature | 450℃ |
Detection limit | ≤1.1pg C/s(n-hexadecane) | ||
Baseline drift (30 min) | ≤2×10⁻13A | ||
Baseline noise | ≤2×10⁻14A | ||
Dynamic Linear Range | ≥10⁷ | ||
Function | Auto-ignition function | ||
Thermal Conductivity Detector (TCD) | Sensitivity | ≥12000mV.mL/mg | |
Baseline drift (30 min) | ≤0.2mV | ||
Baseline noise | ≤0.05mV | ||
Dynamic Linear Range | ≥105 | ||
Function | Air Gap Flow Protection Function | ||
Flame Photometric Detector(FPD) | Maximum operating temperature | 450℃ | |
Detection limit | P≤80fg/s(Methyl parathion)、S≤8pg/s (Methyl parathion) | ||
Dynamic Linear Range | P≥103;S≥102 | ||
Function | Auto-ignition function | ||
Electron Capture Detector (ECD) | radioactive source | 63Ni | |
Maximum operating temperature | 400℃ | ||
Detection limit | ≤1.0×10-14 g/mL(γ-666) | ||
Dynamic Linear Range | ≥104 | ||
Nitrogen-Phosphorus Detector (NPD) | Maximum operating temperature | 400℃ | |
Detection limit | N:≤1×10-12g(N)/s(Azonaphthalene) P:≤5×10-13g(P)/s(Malathion) | ||
Dynamic Linear Range | N≥103、P≥103 |