- Automatic calculation of correlation coefficient (Don’t cover FP-I640)
- Pre-selection of flame sizes
- Flameout protection device
- Measuring range changing
- Multilingual user interface
- Air compressor provide
Flame Photometer, FP-I Series
Description
DESCRIPTION
DESCRIPTION
1. Direct concentration display (Don’t cover FP-I640)
2. Automatic calculation of correlation coefficient (Don’t cover FP-I640)
3. Pre-selection of flame sizes
4. Flameout protection device
5. Measuring range changing
6. Concentration units selectable
7. Multilingual user interface
8. Air compressor provide

Presssure gauge

Power switch

Sample entry chamber

7- inch color touch- screen
SPECIFICATIONS
SPECIFICATIONS
Model | FP-I6450 | FP-I6440 | FP-I6431 | FP-I6430 | FP-I6410 | FP-I640 | |
Operation mode | 7- inch color touch- screen | ||||||
Display value | Concentration value | Optical power | |||||
Data range | 0.000~999.9 | 000.0~999.9 | 0000~9999 | ||||
Testable | K, Na, Li, Ca, Ba | K, Na, Li, Ca | K, Na, Ca | K, Na, Li | K, Na | K, Na | |
Channel Qty. | 5 | 4 | 3 | 2 | |||
Range ppm | K | 0-100 | |||||
Na | 0-160 | ||||||
Li | 0-100 | ○ | 0-100 | ○ | ○ | ||
Ca | 0-1000 | 0-1000 | ○ | ○ | ○ | ||
Ba | 0-3000 | ○ | ○ | ○ | ○ | ○ | |
LOD ppm | K | 0.01 | |||||
Na | 0.01 | ||||||
Li | 0.1 | ○ | 0.1 | ○ | ○ | ||
Ca | 2 | 2 | ○ | ○ | ○ | ||
Ba | 6 | ○ | ○ | ○ | ○ | ○ | |
Linear error | K | 0.195 | |||||
Na | 0.69 | ||||||
Li | 0.15 | ○ | 0.15 | ○ | ○ | ||
Ca | 3 | 3 | ○ | ○ | ○ | ||
Ba | 9 | ○ | ○ | ○ | ○ | ○ | |
Response time | <8s | ||||||
Sample uptake | <6ml/min | ||||||
Stability | < 3% drift over15s when continuously aspirating. | ||||||
Reproducibility | < 3% coefficient of variation for 7 consecutive samples | ||||||
Curve Graph | Display | ○ | ○ | ||||
Printer | Optional build-in thermal printer | ○ | |||||
COM | USB | ○ | |||||
Fuel | LPG | ||||||
Power | AC220V±22V 50Hz±1Hz,250W | ||||||
Packing size | 570mm×530mm×400mm 0.12M3 18kg |