- LN2 Capacity: 5L-500L
- Neck Opening : 40mm
Liquid Nitrogen Storage Container, LNS Series
Description
DESCRIPTION
DESCRIPTION
Product Overview
Liquid nitrogen storage container, LNS Series is mainly used for liquid nitrogen storage. It uses small amount of liquid nitrogen vaporizing to increase the pressure inside the tank, so that the tank can automatically discharge liquid nitrogen to other containers. The stainless steel structure design is suitable for most environment and reduces the rate of evaporation losses. All the models are equipped with pressure building valve, liquid valve, release valve and pressure gauge. All the models are equipped with 4 rollers at the bottom for easy to move. This series liquid nitrogen storage container Mainly applicable to laboratory users and chemical users for liquid nitrogen storage and liquid nitrogen automatic supply.
Product Features
- Unique neck design, low evaporationloss rate;
- A protective operating ring;
- Safe structure;
- Stainless steel tank;
- With rollers for easy to move;
SPECIFICATIONS
SPECIFICATIONS
Model | LNS-5 | LNS-15 | LNS-30 | LNS-50 |
Performance | ||||
LN2 Capacity (L) | 5 | 15 | 30 | 50 |
Neck Opening (mm) | 40 | 40 | 40 | 40 |
Static Evaporation Rate (L/day) | 0.15 | 0.38 | 0.75 | 1 |
Transfusion Volume (L/min) | ≥2 | ≥2 | ≥3 | ≥3 |
Maximum Storage Capacity | ||||
Overall Height (mm) | 652 | 771 | 900 | 1009 |
Outer Diameter (mm) | 354 | 404 | 454 | 506 |
Weight Empty (kg) | 19 | 26 | 38 | 54 |
Weight Full (kg) | 23 | 38.1 | 62.2 | 94.4 |
Standard Working Pressure (mPa) | 0.05 | |||
Maximum Working Pressure (mPa) | 0.09 | |||
Setting Pressure of the First Safety Valve (mPa) | 0.099 | |||
Setting Pressure of the Second Safety Valve (mPa) | 0.15 | |||
Pressure Gauge Indication Range (mPa) | 0~0.25 |
Model | LNS-100 | LNS-150 | LNS-200 | LNS-240 | LNS-300 | LNS-500 |
Performance | ||||||
LN2 Capacity (L) | 100 | 150 | 200 | 240 | 300 | 500 |
Neck Opening (mm) | 40 | 40 | 40 | 40 | 40 | 40 |
Static Evaporation Rate (L/day) | 1.3 | 1.95 | 2.4 | 3 | 3.3 | 5.5 |
Transfusion Volume (L/min) | ≥4 | ≥6 | ≥8 | ≥8 | ≥8 | ≥10 |
Maximum Storage Capacity | ||||||
Overall Height (mm) | 1179 | 1208 | 1288 | 1350 | 1484 | 1607 |
Outer Diameter (mm) | 606 | 706 | 758 | 758 | 858 | 1008 |
Weight Empty (kg) | 86 | 107 | 145 | 148 | 193 | 273 |
Weight Full (kg) | 166.8 | 228 | 307 | / | 435 | 677 |
Standard Working Pressure (mPa) | 0.05 | |||||
Maximum Working Pressure (mPa) | 0.09 | |||||
Setting Pressure of the First Safety Valve (mPa) | 0.099 | |||||
Setting Pressure of the Second Safety Valve (mPa) | 0.15 | |||||
Pressure Gauge Indication Range (mPa) | 0~0.25 |
Static evaporation rate and static holding time is the theoretical value. Actual evaporation rate and holding time will be affected by container usage, atmospheric conditions and manufacturing tolerances.
Series Configuration Guidelines
Conventional Series | K Series |
Standard: Single-layer infusion hose (ZG-DN\10/1.5) Optional: Float level gauge |
Standard: Single-layer infusion hose (ZG-DN10/1.5) Digital display level meter and voltage stabilizing device; Suitable for automatic replenishment system. |
From 100 to 500 liters, K series can be selected, of which LNS-150 and LNS-240 has no conventional series.
Get Social