- Coulometric KF Titration
- 7.0-inch high-contrast color LCD touch screen
- Support oven sampling method, to meet the gas, solid, liquid in the constant moisture, trace moisture measurement.
Karl Fischer Titrator, TITR-50V, TITR-50C
Description
General Features
- Color High contrast LCD touchscreen, 7.0 inches.
- Coulometric KF Titration.
- Support oven sampling method, to meet the gas, solid, liquid in the constant moisture, trace moisture measurement.
TITR-50V: Volumetric Karl Fischer Titration
- Intelligent operation system provides features including user management, method management, sensor management, titrant management, data management and GMP management etc.
- Support Auto Titration Method, KF Titer Determination Method, Oven Titration Method and Oven Blank Titration Method.
- The solvent manager adopts anti-leakage liquid and anti-pour design of waste liquid bottle, supports to fill titrant, drain solution, feed solvent, mix titrant and clean burette, ensure safe handling of Karl Fischer chemicals.
TITR-50C: Coulometric Karl Fischer Titration
- Intelligent operation system provides features including user management, method management, sensor management, data management and GMP management etc.
- Support Auto Titration Method, Oven Titration Method, Oven Blank, and Coefficient correction Method.
- The solvent manager adopts anti-leakage liquid and anti-pour design of waste liquid bottle, supports to drain solution and feed solvent, ensure safe handling of Karl Fischer chemicals.
Data Management
- Auto/manual drift adjust ensures accurate results.
- Selectable units including μg, mg, %, ppm, mg/L and μg/mL.
- User three-level authority management and password management.
- Support for GMP management, the GMP logs achieve the tracking and analysis of the the data.
- Store up to 2000 titration data sets (GLP-compliant).
- Support for RS-232 communication for print and the print format can be alternate.
- Support for USB communication for PC software.
- Support for U disk for method export and import.
- Support for U disk for data export with CSV or PDF format.
- Support for printer, barcode scanner, oven and balance (specific model).
- It needs GMP software to implement GMP management functions.
TITR-50V Includes
- Solvent manager
- 10mL burette
- Volumetric vessel
- Measuring electrod
TITR-50C Includes
- Solvent manager
- Coulometric vessel
- Measuring electrode
- Generator electrode
Specifications
| Model | TITR-50V | TITR-50C |
| Karl Fischer Volumetric Titration | ||
| Water Range | 100μg~250.0mg | / |
| Water Resolution | 1μg | / |
| Water Unit | μg, mg, %, ppm, μg/mL, mg/L | / |
| Water Repeatability | ≤0.3% | / |
| mV Range | (0~2000) mV | / |
| mV Resolution | 0.1mV | / |
| Polarization Current Range | (1~200) μA | / |
| Polarization Current Accuracy | ±3% | / |
| Polarization Current Fluctuation | ±2.5%/30min | / |
| Karl Fischer Coulometric Titration | ||
| Water Range | / | 3.0μg~200mg |
| Water Resolution | / | 0.1μg |
| Water Unit | / | μg, mg, %, ppm, μg/mL, mg/L |
| Water Repeatability | / | ≤0.3% |
| mV Range | / | (0~2000)mV |
| mV Resolution | / | 0.1mV |
| Polarization Current Range | / | (1~200)μA |
| Polarization Current Accuracy | / | ±3% |
| Polarization Current Fluctuation | / | ±2.5%/30min |
| Working Current Accuracy | / | ±0.5% |
| Working Current Fluctuation | / | ±0.2%/10min |
| General | ||
| Power Supply | AC 100~240V, 47~63Hz | AC 100~240V, 47~63Hz |
| Dimensions | 240×370×270mm | 240×370×270mm |
| Net Weight | About 4kg | About 4kg |
| Category | Title | File type/size | Download |
| Brochure | Brochure – Titrator | PDF(3 MB) | download |


Get Social