ECG |
Lead Mode: 5 Leads ( R, L, F, N, C or RA, LA, LL, RL,V) |
Lead selection: I, II, III, avR, avL, avF, V, |
Waveform: 2 ch |
Lead mode: 3 Leads ( R, L, F or RA, LA, LL) |
Lead selection: I, II, III, |
Waveform: 1 ch |
Gain: ′2.5mm/mV, ′5.0mm/mV, ′10mm/mV, ′20mm/mV, auto |
HR and Alarm |
Range Adult: 15 ~ 300 bpm |
Neo/Ped: 15 ~ 350 bpm |
Accuracy : ± 1% or ± 1bpm,which great |
Resolution: 1 bpm |
Sensitivity: > 200 (uV P-P) |
Differential Input Impedance: > 5 MΩ |
CMRR |
Monitor: > 105 dB |
Operation: > 105 dB |
Diagnosis: > 85 dB |
Electrode offset potential: ±300mV |
Leakage Current: < 10 uA |
Baseline Recovery: < 3 S After Defi. |
ECG Signal Range: ±8 m V ( Vp-p ) |
Bandwidth |
Surgery: 1 ~ 15 Hz |
Monitor: 0.5 ~ 35 Hz |
Diagnostic: 0.05 ~ 100 Hz |
Calibration Signal: 1 (mV p-p), Accuracy:±5% |
ST Segment Monitoring Range: Measure and Alarm -2.0 ~ +2.0 mV |
ARR Detecting |
Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, |
TACHY, BRADY, MISSED BEATS, PNP, PNC |
Alarm: Available |
Review: Available |
RESP |
Method: Impedance between R-F(RA-LL) |
Differential Input Impedance: >2.5 MΩ |
Measuring Impedance Range: 0.3~5.0Ω |
Base line Impedance Range: 0 – 2.5 KΩ |
Bandwidth: 0.3 ~ 2.5 Hz |
Resp. Rate |
Measuring and Alarm Range |
Adult: 0 ~ 120 rpm |
Neo/Ped: 0 ~ 150 rpm |
Resolution: 1 rpm |
Accuracy : ±2 rpm |
Apean Alarm: 10 ~ 40 S |
NIBP |
Method: Oscillometric |
Mode: Manual, Auto, STAT |
Measuring Interval in AUTO Mode: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240,480 (Min) |
Measuring Period in STAT Mode: 5 Min |
Pulse Rate Range: 40 ~ 240 bpm |
Alarm Type: SYS, DIA, MEAN |
Measuring and alarm range |
Adult Mode |
SYS: 40 ~ 270 mmHg |
DIA: 10 ~ 215 mmHg |
MEAN: 20 ~ 235 mmHg |
Pediatric Mode |
SYS: 40 ~ 200 mmHg |
DIA: 10 ~ 150 mmHg |
MEAN:20 ~ 165 mmHg |
Neonatal Mode |
SYS: 40 ~ 135 mmHg |
DIA:10 ~ 100 mmHg |
MEAN:20 ~ 110 mmHg |
Resolution Pressure:1mmHg |
Accuracy Pressure Maximum Mean error: ±5mmHg |
Maximum Standard deviation: ±8mmHg |
Overpressure Protection |
Adult Mode: 297±3 mmHg |
Pediatric Mode: 240±3 mmHg |
Neonatal Mode: 147±3 mmHg |
SpO2 |
Measuring Range: 0 ~ 100 % |
Alarm Range: 0 ~ 100 % |
Resolution: 1 % |
Accuracy : 70% ~ 100% ±2 % |
0% ~ 69% unspecified |
Actualization interval: about 1 Sec. |
Alarm Delay: 10 Sec. |
Pulse Rate |
Measuring and Alarm Range: 20~300bpm |
Resolution: 1bpm |
Accuracy : ±2bpm |
TEMPERATURE |
Channel: 2 |
Measuring and Alarm Range: 0 ~ 50 °C |
Resolution: 0.1°C |
Accuracy : ±0.1°C |
Average Time Constant:<10 Sec. |
IBP |
Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2 |
ART: 0 ~ 300 mmHg |
PA: -6 ~ 120 mmHg |
CVP/RAP/LAP/ICP: -10 ~ 40 mmHg |
P1/P2: -10 ~ 300 mmHg |
Press Sensor |
Sensitivity: 5 uV/V/mmHg |
Impedance: 300-3000Ω |
Resolution: 1 mmHg |
Accuracy : ±2% or ±1mmHg, which great |
Actualization interval: about 1 Sec |
Standard: ECG, Respiration, NIBP, SpO2, Pulse Rate, Temperature |
Optional: Nellcor SpO2, EtCO2,IBP, Touch screen, Thermal Recorder,Wall mount, Trolley, Central station,VGA |