- Temperature Range: 0~60 ℃
- Display Resolution: 0.1 ℃
- Temperature Stability: Heating: ±0.5℃; Cooling: ±1℃
Cooling Incubator, ICB-BII Series
Description
DESCRIPTION
DESCRIPTION
Applications
Biochemical incubators are suitable for scientific research and production departments such as environmental protection, medical treatment, sanitation, epidemic prevention, drug inspection, agriculture, livestock and aquaculture. ..
Features
Energy-saving
- Adopts new synthetic silicone sealing, better sealing effect, preventing heat loss effectively.
- Automatically adjust the control parameters according to the ambient temperature and load to achieve the optimal heating method and reduce energy consumption.
- High efficiency and energy saving compressor, stable temperature control and low power consumption.
Intelligent control
- Real-time parameter high efinition LCD display,intelligent microcomputer control.
- Simple and convenient operation, can quickly set temperature and time parameters.
- Standard RS-485 communication interface
Internal circulation system
- High-quality compressor, automatic control for hot and cold.
- Back suction forced convection system.
- Adopts high-quality fan and large impeller fan, has higher ventilation efficiency and better temperature uniformity.
Dedicated sheet metal processing
- The shelf can be disassembled and can be freely adjusted up and down.
- The inner chamber is made of 304 high-quality mirror stainless steel, with strong corrosion resistance.
Humanized design
- Equipped with lamp inside, easy observation.
- Equipped with power supply outlet inside chamber, easily provide electricity for small instruments.
- The surface coated, good appearance and maintenance.
- Double door design, the inner door is made of tempered glass for easy observation,magnetic sealing design for outer door, good sealing.
Safety
- Equipped with leakage protection device to ensure the safety of operators and other equipment in the laboratory.
- Equipped with spare temperature control which ensures the product work normally even the main temperature control failed(for heating)
- Over-temperature control sound alarm system
Optional:
- Multi-segment programmable control
- Φ25mm (50mm) test hole
- Spare shelf
- USB port
- Printer
SPECIFICATIONS
SPECIFICATIONS
Model | ICB-B80 II | ICB-B150 II | ICB-B250 II | ICB-B300 II | ICB-B400 II | |
Capacity(L) | 80 | 150 | 250 | 300 | 400 | |
Circulation Mode | Forced convection | |||||
Dimension(W*D*H) | Internal (cm) | 40*37*55 | 45*42*85 | 48*49*107 | 52*50*117 | 58*54*127 |
External (cm) | 54*59*107 | 59*64*137 | 62*70*159 | 66*71*169 | 72*75*179 | |
Temperature | Range (℃) | 0~60 | ||||
Fluctuation (℃) | High Temp.:±0.5 Low Temp.:±1 | |||||
Resolution (℃) | 0.1 | |||||
Uniformity (℃) | ±1 | |||||
Sensor | PT100 | |||||
Controller | Microcomputer control | |||||
Display | High efinition LCD | |||||
Timer(min) | 1~9999 | |||||
Material | Internal | Mirror 304 stainless steel | ||||
External | Coated cold rolled steel | |||||
Shelves(Standard / Max ) | 2 / 8 | 3 / 13 | 3 / 16 | 3 / 18 | 3 / 20 | |
Test Hole | Φ25mm (50mm) | |||||
Refrigerant | R134a (CFC free) | |||||
Safety Device | Over temperaure Visible& Audible alarm,digital independent temp. limiter,leakage protector | |||||
Electricity | Voltage/Frequency | AC 220/110V, 50/60Hz | ||||
Consumption (W) | 180 | 250 | 350 | 400 | 550 | |
N.W./G.W. (kg) | 75/105 | 85/122 | 159/189 | 163/195 | 145/200.5 | |
Package Dimension(W*D*H) (cm) | 80*64*127 | 71*76*157 | 76*79*182 | 79*79*186 | 86*86*201 | |
Optional | Multi-segment programmable control Φ25mm (50mm) test hole USB port Printer spare shelf |