Chat with us, powered by LiveChat

Fully Automatic Enclosed Tissue Processor, TSP-CV1

  1. Enclosed Tissue treatment system, vacuum type
  2. Number of Cups: 12
  3. Capacity of Single Cup: 6000ml
  4. Processing capacity: Max 300 pcs/cassette
  5. 10 Set of Programmers stored in the system
  6. Temperature Range: solvent:≤45℃, Wax: 58℃-70℃

Share the product

Category:

Description

  1. Operator Interface of 15 inch color LCD touch screen
  2. Enclosed Tissue treatment system with no pollution of gas leakage, to meet requirement of environmental protection.
  3. Mode of Sample processing: Sample not move, Reagent
  4. It can avoid the risk of powerfailure of machine, or other mechanical failure.
  5. Timing Mode: Finishing time of tissue processing any day in the week can be set
  6. Function of Protection System of Power Once main power restored, the previous protocols is automatically resumed and continue to work in order.
  7. Special design of Reagent Station makes the liquid totally back-flow to avoid any mixture and to extend using time of reagent.
  8. Patented design of cassette holder makes reagent and samples thoroughly contact with each other, itimproves effection of tissue processing and reduce working time.
  9. Dual Purpose of processing: one key switch normal processing motion and Rapid processing motion
  10. Rapid Processing motion stirred in day time finished in 3 hours, NormalProcessing Motion stirred in the evening.
  11. Device for Air pressure transfer: Formal air pressure transfer of several solenoid valves replaced by one single mechanical structure replace formal in order to avoid high fault error.
  12. Functionfor changing wax automatically:
  13. Wax in No.1 Container wash into waste container,then, wax in NO.2 container goes into No.1 Container,then,No.3 goes into No.2,and so on
  14. Functionof over temperature protection occurred by accidental heating
  15. Functionfor magnetic stirring which shorten time of tissue processing and make good processing effection.
  16. Alarm and Screen display when finishing work.
  17. Capacity of tissue processing: 300pcs cassette at most.

Specifications

Numbers of Reagent Station 12
Tissue Processing Station  NO.1 to No.9 Station
Washing Station  No.14 to No.16 Station
Numbers of Wax Tank  3(No.10,12 and 13)
Numbers of Working Station 1
Capacity of Processing Station  9Liters
Capacity of Reagent Station 6L
Single wax tank capacity 6L
Dehydration basket capacity 163*163*198 (mm) (300 dehydration cassettes)
Temperature of Processing Station:
Medium is the solvent RT – 45℃
Medium is the Wax  58℃-80℃
Operation pressure <0.1Mpa
Single reagent processing time Tanks 1 & 2: 0~6000 minutes, others: 0~300 minutes
Programmable settings 10 sets.
Wax melt time Not more than 3 hours
Power input <1500VA+10%
Working Voltage 220VAC±10%,50/60HZ
Time for Inflow Liquid  no more than 5Min
Time for Exclude Liquid  no more than 5Min
Stirring time 0~9 minutes
Interval time for Stirring 0~12 minutes
Operation mode Manual and automatic
Overall Dimension 650×650×1350mm(W×D×H)
Net Weight 115kg
Package Dimension (W*D*H)(mm) 770*770*1480
CategoryTitleFile type/sizeDownload
BrochureBrochure – Tissue ProcessorPDF(377 KB)download

    Get Quotes Now!

    We will usually contact you within 24 hours. You could also call us (+86-531-88982330) during working hours ( 8:30 am to 5:45 pm UTC+8 Mon.~Sat. ) or use the website live chat to get prompt reply.

    Contact Info

    Rm. 2014, Bldg. 3, Ligaoguojihuayuan, No. 1222, West Aoti Road, Lixia District, Jinan, Shandong

    Phone: +86-531-88982330

    Fax: +86-531-88983691

    Web: www.infitek.com

    Go to Top