- Enclosed Tissue treatment system, vacuum type
- Number of Cups: 12
- Capacity of Single Cup: 6000ml
- Processing capacity: Max 300 pcs/cassette
- 10 Set of Programmers stored in the system
- Temperature Range: solvent:≤45℃, Wax: 58℃-70℃
Fully Automatic Enclosed Tissue Processor, TSP-CV1
Description
- Operator Interface of 15 inch color LCD touch screen
- Enclosed Tissue treatment system with no pollution of gas leakage, to meet requirement of environmental protection.
- Mode of Sample processing: Sample not move, Reagent move.
- It can avoid the risk of power failure of machine, or other mechanical failure.
- Timing Mode: Finishing time of tissue processing any day in the week can be set
- Function of Protection System of Power Failure. Once main power restored, the previous protocols is automatically resumed and continue to work in order.
- 10 Set of Programmers stored in the system and can be set for purpose.
- Special design of Reagent Station makes the liquid totally back-flow to avoid any mixture and to extend using time of reagent.
- Patented design of cassette holder makes reagent and samples thoroughly contact with each other, it improves effection of tissue processing and reduce working time.
- Dual Purpose of processing: one key switch normal processing motion and Rapid processing motion
- Rapid Processing motion stirred in day time finished in 3 hours, Normal Processing Motion stirred in the evening.
- Device for Air pressure transfer: Formal air pressure transfer of several solenoid valves replaced by one single mechanical structure replace formal in order to avoid high fault error.
- Function for changing wax automatically:Wax in No.1 Container wash into waste container, then, wax in NO.2 container goes into No.1 Container,then,No.3 goes into No.2,and so on
- Function of over temperature protection occurred by accidental heating
- Function for magnetic stirring which shorten time of tissue processing and make good processing effection.
- Alarm and Screen display when finishing work.
- Capacity of tissue processing: 300 pcs cassette at most.
Specifications
Model | TSP-CV1 |
Numbers of Reagent Station | 12 |
Tissue Processing Station | NO.1 to No.9 Station |
Washing Station | No.14 to No.16 Station |
Numbers of Wax Tank | 3(No.10,12 and 13) |
Numbers of Working Station | 1 |
Capacity of Processing Station | 9L |
Capacity of Reagent Station | 6L |
Temperature of Processing Station | Medium is the solvent: ≤45℃ |
Medium is the Wax: 58℃-70℃ | |
Operation pressure | <0.1Mpa |
Wax temperature | 58℃-70℃ |
Wax melt time | <3 hours |
Power input | <1500VA+10% |
Time for immerse | 0 – 99 hours 59 minutes |
Working Voltage | 220VAC±10%,50/60HZ |
Time for Inflow Liquid | no more than 5Min |
Time for Exclude Liquid | no more than 5Min |
Stirring time | set randomly |
Interval time for Stirring | set randomly |
Operation mode | manual and automatic |
Overall Dimension | 650×650×1350mm(W×D×H) |
Net Weight | 145kg |